EasyManua.ls Logo

Endress+Hauser cerabar S - Page 96

Endress+Hauser cerabar S
172 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Description of parameters (ToF Tool) Cerabar S/Deltabar S
96 Endress+Hauser
P01-xMx7xxxx-19-xx-xx-xx-028
Fig. 20: Quick Setup menu, "Flow" measuring mode (Differential pressure transmitter Deltabar S only)
DAMPING VALUE
Entry
Enter damping time (time constant τ).
The damping affects the speed at which all subsequent elements, such as the on-site
display, measured value and OUT value of the Analog Input Block react to a change in the
pressure.
Input range:
0.0...999.0 s
Factory setting:
2.0 s or as per order specifications
ToF Tool, Table 2: QUICK SETUP "Level"
Parameter name Description
OPERATING MENU
MAX FLOW
MAX. PRESS. FLOW
POS. ZERO ADJUST
DAMPING VALUE
Flow
Pressure
Level
MEASURED VALUE
QUICK SETUP
MEASURING MODE
ToF Tool, Table 3: QUICK SETUP "Flow"
Parameter name Description
This menu displays the most important parameters for the "Flow" measuring mode.
Prerequisite:
Deltabar S differential pressure transmitter
•MEASURING MODE = flow
Note:
See also
Page 125, ToF Tool, table 10: BASIC SETUP
Page 130, ToF Tool, table 13: EXTENDED SETUP
Page 134, ToF Tool, table 15: TOTALIZER SETUP
Page 89 ff, Section 9 "Flow measurement (ToF Tool)".
MEASURING MODE
Selection
Select the measuring mode.
The operating menu is structured according to the selected measuring mode.
!
Note!
When the measuring mode is changed, no conversion takes place. The device has to be
recalibrated if the measuring mode is changed. See also Page 89, Section 9 "Flow
measurement (ToF Tool)".
Options:
Pressure
Level
Deltabar S: flow
Factory setting:
Pressure

Related product manuals