Gammapilot M FMG60 with FOUNDATION Fieldbus Operation
Endress+Hauser 33
Function of the keys
Key(s) Meaning
O
or
V
Navigate upwards in the selection list
Edit numeric value within a function
S
or
W
Navigate downwards in the selection list
Edit numeric value within a function
X
or
Z
Navigate to the left within a function group
F
Navigate to the right within a function group, confirmation.
O
and
F
or
S
and
F
Contrast settings of the LCD
O
and
S
and
F
Hardware lock / unlock
After a hardware lock, operation of the device via display or communication is not possible!
The hardware can only be unlocked via the display. An unlock parameter must be entered
to do so.