Operation Gammapilot M FMG60 with FOUNDATION Fieldbus
34 Endress+Hauser
5.2.2 The operating menu
Function code
The functions of the Gammapilot M are arranged in an operating menu. To ensure easy orientation
within the menu, a unique position code is indicated on the display for each function. This code
consists of one alphabetic and two numeric characters.
L00-FMG60xxx-07-00-00-en-002
• The alphabetic character specifies the current measuring mode of the Gammapilot M:
– L: level
– S: limit (switch)
– D: density
– C: concentration
– *: no measuring mode selected yet
• The first numeric character identifies the function group:
– basic setup *0
– calibration *1
– safety settings *2
– ...
• The second numeric character numbers the individual functions within the function group:
basic setup *0
– today's date *01
– beam type *02
– isotope *03
– operating mode *04
– ...
Hereafter, the position is always given in brackets after the function name. "*" (not yet selected) is
always indicated as the measurement method, e.g. "present date" (*01).
measuring mode
functiongroup
function