Gammapilot M FMG60 with FOUNDATION Fieldbus Operation
Endress+Hauser 35
Operation using the onsite display VU 331
L00-FMG60xxx-19-00-00-en-006
1. Change from Measured Value Display to Group Selection by pressing F.
2. Press S or O to select the required Function Group and confirm by pressing F.
The active selection is marked by a ✓ in front of the menu text.
3. Activate Edit mode with O or S .
Selection menus
a. Select the required parameter in the function selected with S or O.
b. F confirms selection; ✓appears in front of the selected parameter.
c. F confirms the edited value; system quits edit mode.
d. O and S (= Q) interrupts selection; system quits edit mode.
Typing in numerals and text
a. Press O or S to edit the first character of the numeral / text.
b. F positions the cursor at the next character; continue with a. until you have completed
your input.
c. If a ↵ symbol appears at the cursor, press F to accept the value entered; system quits edit
mode.
d. If a ← symbol appears at the cursor, press F to return to the previous character (e.g. for
correction of entries).
e. O and S (= Q) interrupts selection; system quits edit mode.
4. Press F to select the next function.
5. Press O and S (= Q) once; return to previous function.
Press O and S (= Q) twice; return to Group Selection.
6. Press O and S (= Q) to return to Measured value display.
X
X
X
X
S
S
O
O
FF
>3 s
F
...
2x
ENDRESS + HAUSER
E
+
–
...
137Cs
60Co
no compens.
return to
group selection
basic setup
calibration
linearisation
temperature compensation
safety setting
todays date
isotope