Operation Gammapilot M FMG60 with FOUNDATION Fieldbus
40 Endress+Hauser
5.6.2 Hardware settings
Two DIP switches in connection compartment B allow the hardware setting for write protection
and the simulation function. A diagram in the connection compartment explains the position of the
switches (ON and OFF)
L00-FMxxxxxx-19-00-00-en-002
Device identification
Foundation Fieldbus identifies the device by its identification code (device ID) and automatically
allocates an appropriate field address. There is no separate hardware switch for this purpose.
Switch Position Meaning
SIM ON (down) Simulation possible in the configuration tool.
OFF (up) Simulation not possible in the configuration tool.
WP
(write protection)
ON (down) Parameters can only be read.
OFF (up) Parameters can be written and read.
IN CASCADE OUT
OFF
ON
+ --+
B
SIM WP
OFF
ON