Operation Gammapilot M FMG60 with FOUNDATION Fieldbus
46 Endress+Hauser
Parameter list of the Gammapilot M Transducer block
Parameter Position
Indicator
rel.
Index
Variable Name Size
[bytes]
Type Read Write Storage
Class
0 EH_USONICLEVEL_CAL_BASIC
Standard parameters
ST_REV 1 ST_REV 2 UNSIGNED16 X static
TAG_DESC 2 TAG_DESC 32 OCTET_STRING X X static
STRATEGY 3 STRATEGY 2 UNSIGNED16 X X static
ALERT_KEY 4 ALERT_KEY 1 UNSIGNED8 X X static
MODE_BLK 5 MODE_BLK 4 DS-69 X X static
BLOCK_ERROR 6 BLOCK_ERROR 2 BIT_STRING X dynamic
UPDATE_EVT 7 UPDATE_EVT 14 DS-73 X dynamic
BLOCK_ALM 8 BLOCK_ALM 13 DS-72 X X dynamic
Transducer Directory Entry 9 TRANSDUCER_DIRECTORY 2 UNSIGNED16 X non-vol.
Transducer Type 10 TRANSDUCER_TYPE 2 UNSIGNED16 X non-vol.
Transducer Error 11 XD_ERROR 1 UNSIGNED8 X static
Collection Directory 12 COLLECTION_DIRECTORY 4 UNSIGNED32 X non-vol.
Primary Value Type 13 PRIMARY_VALUE_TYPE 2 UNSIGNED16 X X static
Primary Value 14 PRIMARY_VALUE 5 DS-65 X dynamic
Primary Value Range 15 PRIMARY_VALUE_RANGE 11 DS-68 X non-vol.
Secondary Value 16 SECONDARY_VALUE 5 DS-65 X dynamic
Secondary Value Unit 17 SECONDARY_VALUE_UNIT 2 UNSIGNED16 X X static
Third Value 18 THIRD_VALUE 5 DS-65 X dynamic
Third Value Unit 19 THIRD_VALUE_UNIT 2 UNSIGNED16 X X static
Instrument specific parameters
measured value *00 20 SENSOROUTVALUE 4 FLOAT X dynamic
Present date day *01 21 PRESENTDATEDAY 1 UNSIGNED8 X X dynamic
Present date month *01 22 PRESENTDATEMONTH 1 UNSIGNED8 X X dynamic
Present date year *01 23 PRESENTDATEYEAR 1 UNSIGNED8 X X dynamic
Present date hour *01 24 PRESENTDATEHOUR 1 UNSIGNED8 X X dynamic
Present date minute *01 25 PRESENTDATEMINUTE 1 UNSIGNED8 X X dynamic
Ray type *02 26 RAYTYPE 1 ENUM8 X X non-vol.
Radiation source *03 27 RADIATIONSOURCE 1 ENUM8 X X non-vol.
Master slave mode *04 28 MASTERSLAVEMODE 1 ENUM8 X X non-vol.
Measurement method *05 29 MEASUREMENTMODE 1 ENUM8 X X non-vol.
Density unit *06 30 DENSITYUNIT16 2 ENUM16 X X non-vol.
Min. density *07 31 DENSITY4MAVALUE 4 FLOAT X X non-vol.
Max. density *08 32 DENSITY20MAVALUE 4 FLOAT X X non-vol.
Pipe diameter unit *09 33 PIPEDIAMUNIT16 2 ENUM16 X X non-vol.
Pipe diameter *0A 34 PIPEDIAMETER 4 FLOAT X X non-vol.
Output damping *0B 35 INTEGRATIONTIME 2 UNSIGNED16 X X non-vol.
Background calibration *10 36 BACKGRCALIB 1 ENUM8 X X non-vol.
Avg. pulse rate *11 37 AVGPULSERATE 4 INTEGER32 X dynamic
Background pulse rate *12 38 BACKGRPULSERATE 4 INTEGER32 X X non-vol.