Commissioning Gammapilot M FMG60 with FOUNDATION Fieldbus
60 Endress+Hauser
"Empty" or "free" calibration
(for details see Section 6.4.3, P. 74)
Step Function Action/comments
1 • The radiation remains switched on
•Empty tank
Level: max. 40%
Level limit: beam path free
2 Calibration point level (*13) Select "empty/free" with O or S;
confirm with F
The following function for level measurement only
3 Value empty (*17) Enter the current level (0% - 40%)
The following functions for level and level limit measurement
4 Calibration start stop (*15) Select "start" with O or S;
confirm with F
5 Avg. pulse rate (*11) Wait until the value is stable, then confirm with F
6 Calibration start stop (*15) Select "stop/edit" with O or S;
confirm with F
7 Empty calibration (*16) Confirm value with F
8 Calibration next point (*19) Select "no" with O or S;
Confirm with F
This completes the calibration.