EasyManuals Logo

Endress+Hauser Liquiline CA80SI User Manual

Endress+Hauser Liquiline CA80SI
162 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #119 background imageLoading...
Page #119 background image
Liquiline System CA80SI Diagnostics and troubleshooting
Endress+Hauser 119
No. Message Factory settings Tests or remedial action
S
1)
D
2)
F
3)
970 Curr. input overload S On On Current input overloaded
The current input is switched off from 23 mA due
to overload and reactivated automatically when a
normal load is present.
971 Current Input low S On On Current input too low
At 4 to 20 mA, the input current is less than the
lower failure current.
Check the input for short-circuiting
972 Curr. input > 20 mA S On On Current output range exceeded
973 Current Input < 4 mA S On On Current output range undershot
974 Diagnostics confirm. C Off Off User has acknowledged the message displayed in
the measuring menu.
975 Device restart C Off Off Device reset
976 PFM value high S On Off Pulse frequency modulation: output signal
exceeded/undershot. Measured value outside the
specified range.
Sensor in air
Air pockets in assembly
Incorrect flow to sensor
Sensor fouled
1. Clean sensor
2. Check plausibility
3. Adjust the PFM configuration
977 PFM value low S On Off
978 ChemoClean Failsafe S On On No feedback signal detected within the configured
period.
1. Check the application
2. Check the wiring
3. Extend the duration
990 Deviation limit F On On Redundancy: limit value of percentage deviation
exceeded
991 CO2 conc. range F On On CO
2
concentration (degassed conductivity) outside
the measuring range
992 pH calculation range F On On pH calculation outside the measuring range
993 rH calculation range F On On rH calculation outside the measuring range
994 Difference conduct. F On On Dual conductivity outside the measuring range
995 Mathematical error S On On Incorrect calculation result
1. Check the mathematical functions.
2. Check the input variables.
1) Status signal
2) Diagnostic message
3) Error current
12.6.2 Sensor-specific diagnostic messages
Operating Instructions "Memosens", BA01245C
12.7 Pending diagnostic messages
The Diagnostics menu contains all the information on the device status. Furthermore,
various service functions are available.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Endress+Hauser Liquiline CA80SI and is the answer not in the manual?

Endress+Hauser Liquiline CA80SI Specifications

General IconGeneral
Power Supply24 V DC
Measured VariablesConductivity
Measuring Range0 to 200 mS/cm
Measurement Range (General)0.0005 to 5 mg/l SiO₂
Accuracy±1% of measured value
Accuracy (General)±1% of measured value
Repeatability±0.5% of measured value
ResolutionpH: 0.01, Conductivity: 0.001 μS/cm, DO: 0.01 mg/L, ORP: 1 mV
Ambient Temperature-20°C to +60°C
Operating Temperature-20°C to +60°C
Output Signal4 to 20 mA
CommunicationHART, PROFIBUS PA, Modbus, Ethernet/IP
Communication ProtocolHART, PROFIBUS PA, Modbus, Ethernet/IP
ProtectionIP66
Housing MaterialPolycarbonate

Related product manuals