Technical data Liquiline System CA80SI
148 Endress+Hauser
16.5 Protocol-specific data
Modbus RS485
Protocol RTU/ASCII
Function codes 03, 04, 06, 08, 16, 23
Broadcast support for function codes 06, 16, 23
Output data 16 measured values (value, unit, status), 8 digital values (value,
status)
Input data 4 setpoints (value, unit, status), 8 digital values (value, status),
diagnostic information
Supported features Address can be configured using switch or software
Modbus TCP
TCP port 502
TCP connections 3
Protocol TCP
Function codes 03, 04, 06, 08, 16, 23
Broadcast support for function codes 06, 16, 23
Output data 16 measured values (value, unit, status), 8 digital values (value,
status)
Input data 4 setpoints (value, unit, status), 8 digital values (value, status),
diagnostic information
Supported features Address can be configured using DHCP or software
Web server The Web server enables full access to the device configuration, measured values,
diagnostic messages, logbooks and service data via standard WiFi/WLAN/LAN/GSM or 3G
routers with a user-defined IP address.
TCP port 80
Supported features • Remote-controlled device configuration(1 session)
• Save/restore device configuration (via SD card)
• Logbook export (file formats: CSV, FDM)
• Access to Web server via DTM or Internet Explorer
• Login
• Web server can be switched off
EtherNet/IP
Log EtherNet/IP
ODVA certification Yes
Device profile Generic device (product type: 0x2B)
Manufacturer ID 0x049E
h
Device type ID 0x109F
Polarity Auto-MIDI-X
Connections CIP 12
I/O 6
Explicit message 6
Multicast 3 consumers
Minimum RPI 100 ms (default)
Maximum RPI 10000 ms
System integration EtherNet/IP EDS