Diagnostics and troubleshooting Liquiline System CA80SI
120 Endress+Hauser
The following messages are directly displayed every time you enter the menu:
• Most important message
Diagnostic message recorded with the highest criticality level
• Past message
Diagnostic message whose cause is no longer present.
All the other functions in the Diagnostics menu are described in the following chapters.
If diagnostic message M313 Liquidsensor appears five times in succession when
running a program, the active program is aborted for reasons of safety. This behavior
on the part of the device cannot be altered by deactivating the diagnostic message
under Menu/Setup/General settings/Extended setup/Diagnostics settings/Diag.
behavior .
12.8 Diagnostics list
All the current diagnostic messages are listed here.
A time stamp is available for each message. Furthermore, the user also sees the
configuration and description of the message as saved in Menu/Setup/General settings/
Extended setup/Diagnostics settings/Diag. behavior .
12.9 Logbooks
12.9.1 Available logbooks
Types of logbooks
• Logbooks physically available (all apart from the overall logbook)
• Database view of all logbooks (=overall logbook)
Logbook Visible in Max.
entries
Can be
disabled
1
)
Logbook
can be
deleted
Entries
can be
deleted
Can be
exported
Overall logbook All events 20000 Yes No Yes No
Calibration logbook Calibration events 75 (Yes) No Yes Yes
Operation logbook Configuration events 250 (Yes) No Yes Yes
Diagnostics logbook Diagnostic events 250 (Yes) No Yes Yes
Analyzer event
logbook
Analyzer events 19500
2)
No No Yes Yes
Analyzer calibration
logbook
Analyzer calibration
logbook
250 (Yes) No Yes Yes
Analyzer data logbook Analyzer data
logbooks
20000
2)
No No Yes Yes
Analyzer absorbance
data logbook
Absorption data
logbook
5000 No No Yes Yes
Analyzer raw values
logbook
Raw data logbook 5000 No No Yes Yes
Version logbook All events 50 No No No Yes
Hardware version
logbook
All events 125 No No No Yes