Technical data Liquiline System CA80SI
156 Endress+Hauser
16.8 Environment
Ambient temperature +5 to +40 °C (41 to 104 °F)
Storage temperature -20 to +60 °C (-4 to 140 °F)
Humidity 10 to 95 %, non-condensating
Degree of protection IP55 (cabinet, analyzer stand), TYPE 3R (cabinet, analyzer stand)
Electromagnetic
compatibility
Interference emission and interference immunity as per EN 61326-1:2013, Class A for
Industry
Electrical safety According to EN/IEC 61010-1:2010, Class I equipment
Low voltage: overvoltage category II
For installations up to 2000 m (6500 ft) above MSL
Degree of contamination The product is suitable for pollution degree 2.
16.9 Process
Sample temperature +5 to +45 °C (41 to 113 °F)
Process pressure 1 to 5 bar (14.5 to 72.5 psi)
Recommended:1.5 to 3 bar (21.8 to 43.5 psi)
Sample flow rate 60 to 200 ml/min
Recommended:70 ml/min
Consistency of the sample Particle-free
Process connection Plug-in hose connector for rigid hoses with OD 6 mm
16.10 Mechanical construction
Dimensions →  17