Electrical connection Liquiline System CA80SI
36 Endress+Hauser
DIP switches on front of module
DIP Factory
setting
Assignment
1-128 ON Bus address (→ "Commissioning/communication")
OFF Write protection: "ON" = configuration not possible via the bus, only via local
operation
Service OFF If the switch is set to "ON" , the user settings for Ethernet addressing are saved
and connection settings programmed into the device at the factory are activated:
IP address=192.168.1.212, Subnet mask=255.255.255.0, Gateway=0.0.0.0,
DHCP=Off.
If the switch is set to "OFF" , the saved user settings are reactivated.
7.4 Hardware settings
7.4.1 Bus termination (module 485 only)
There are 2 ways to terminate the bus:
1. Internal termination (via DIP switch on module board)
"ON"
"OFF"
 50 DIP switch for internal termination
Using a suitable tool such as a tweezer, move all four DIP switches to the "ON" position.
The internal termination is used.
390 W
220 W
390 W
VP
DGND
B/B' A/A'
 51 Structure of internal termination
2. External termination
Leave the DIP switches on the module board in the "OFF" position (factory setting).
Connect the external termination to terminals 81 and 82 on the front of module 485
for 5V power supply.
The external termination is used.