Liquiline System CA80SI Electrical connection
Endress+Hauser 35
DIP switches on front of module
DIP Factory setting Assignment
1-128 ON Bus address (→ "Commissioning/communication")
OFF Write protection: "ON" = configuration not possible via the bus, only via local
operation
Service OFF If the switch is set to "ON" , the user settings for Ethernet addressing are saved
and connection settings programmed into the device at the factory are
activated: IP address=192.168.1.212, Subnet mask=255.255.255.0,
Gateway=0.0.0.0, DHCP=Off.
If the switch is set to "OFF" , the saved user settings are reactivated.
Module ETH
48 Module
Ethernet
DIP switch
on
1
2
4
8
16
32
64
128/SW
Service
COM
SF
BF
PWR
49 Wiring
diagram
LEDs on front of module
LED Designation Color Description
RJ45 LNK/ACT GN • Off = Connection is not active
• On = Connection is active
• Flashing = Data transmission
RJ45 10/100 YE • Off = Transmission rate 10 MBit/s
• On = Transmission rate 100 MBit/s
PWR Power GN Supply voltage is applied and module is initialized
BF Bus failure RD Not used
SF System failure RD Device error
COM Communication YE Modbus message sent or received