Maintenance Liquiline System CA80SI
132 Endress+Hauser
Interval Maintenance work
Every 12 months Replace cuvette
Clean magnetic stir bar
Check hose on hose pump for wear, and replace if necessary
Replace hoses
Replace filter mats
If required Cleaning the housing
Clean cuvette
Replace cuvette
Clean magnetic stir bar
Rinse the system
Replace hose connector
Clean the filter (backwash)
13.2 Cleaning
L
CAUTION
Risk of injury from reagents escaping
Clean the system every time before replacing consumables.
13.2.1 Cleaning the housing
You can find an overview of the maintenance intervals here: →  131.
Clean the front of the housing using commercially available cleaning agents only.
The front of the housing is resistant to the following in accordance with DIN 42 115:
Ethanol (for a short time)
Diluted acids (max. 2% HCl)
Diluted bases (max. 3% NaOH)
Soap-based household cleaning agents
NOTICE
Cleaning agents not permitted
Damage to the housing surface or housing seal
Never use concentrated mineral acids or alkaline solutions for cleaning.
Never use organic cleaners such as acetone, benzyl alcohol, methanol, methylene
chloride, xylene or concentrated glycerol cleaner.
Never use high-pressure steam for cleaning.
13.2.2 Rinsing the system
You can find an overview of the maintenance intervals here: →  131.
13.3 Replacing reagents
You can find an overview of the maintenance intervals here: →  131.
Removing bottles
1. Open the analyzer.
2. Stop the sample supply.
3. Press the soft key MODE and select Manual mode item.
The display shows Current mode- Manual.
Wait until all operations have stopped. Ongoing operations can be stopped under
Menu/Operation/Manual operation.