Liquiline System CA80SI Installation
Endress+Hauser 25
4-/6-channel device: Installation area for sample channel switching and panel with
pressure relief valves and filters
H
D
< 1 m
(3.28 !)
< 1 m
(3.28 !)
D + H < 3m
(9.84 !)
L < 1m
(3.28 !)
A0036574
 25 Permitted installation area, can be installed to the left or right of the analyzer, engineering unit m (ft)
H
D
< 1 m
(3.28 !)
< 1 m
(3.28 !)
D + H < 3m
(9.84 !)
< 1 m
(3.28 !)
A0036667
 26 Permitted installation area, engineering unit m (ft)
6.6 4-/6-channel version: Installing panel with sample
channel switching
Mounting materials are not supplied.
The mounting materials must be provided by the customer onsite.
Panel dimensions