Liquiline System CA80SI System integration
Endress+Hauser 39
8 System integration
8.1 Web server
8.1.1 Connection
Connect the computer's communication cable to the RJ45 port of module 485 or
module ETH.
485 / ETH
Web Browser
A0026779
 55 Web server/Ethernet connection
8.1.2 Establishing the data connection
To ensure that your device has a valid IP address, you must disable the DHCP parameter in
the Ethernet settings. (Menu/Setup/General settings/Extended setup/Ethernet/
Settings)
You can then set the IP address in the same menu.
1. Start your PC.
2. First, configure a manual IP address in the network connection settings of the
operating system.
Example: Microsoft Windows 7
3. In the Control Panel, go to the Network and Sharing Center. You should see an active
"Local Area Connection" link.
4. Select the "Local Area Connection" link.
5. In the pop-up window select the "Properties" button.
6. Double-click "Internet Protocol Version 4 (TCP/IPv4)".
7. Select "Use the following IP Address".
8. Enter the desired IP address. This address must be in the same subnetwork as the IP
address of the device, e.g.:
IP address for Liquiline: 192.168.1.212 (as configured previously)
IP address for PC: 192.168.1.213.
9. Start the internet browser.
10. If you use a proxy server to connect to the Internet:
Disable the proxy (browser settings under "Connections/LAN settings").
11. Enter the IP address of your device in the address bar (192.168.1.212 in the
example).
The system takes a few moments to establish the connection and then the CM44
web server starts. You might be asked for a password. The factory setting is
"admin" for the user name and "admin" for the password.