Diagnostics and troubleshooting Liquiline System CA80SI
126 Endress+Hauser
DIAG/System information
Function Options Info
System modules
Backplane Read only
Description
Serial number
Order code
Hardware version
Software version
This information is provided for every electronics
module available. Specify the serial numbers and
order codes  when servicing, for example.
Base
Display module
Extension module 1 ... 8
Sensors Read only
Description
Serial number
Order code
Hardware version
Software version
This information is provided for every sensor
available. Specify the serial numbers and order
codes when servicing,  for example.
Save system information
Save to SD card File name assigned
automatically (includes a
time stamp)
The information is saved on the SD card in a
"sysinfo" subfolder.
The csv file can be read and edited in MS Excel, 
for example. This file can be used when servicing
the device.
1) Provided you give the manufacturer all the information about changes to the hardware.
12.11 Sensor information
Select the channel you want from the list of channels.
Information in the following categories is displayed:
Extreme values
Extreme conditions to which the sensor was previously exposed, e. g. min./max.
temperatures
2)
Operating time
Operating time of the sensor under defined extreme conditions
Calibration information
Calibration data of the last calibration
Sensor specifications
Measuring range limits for main measured value and temperature
General information
Information on sensor identification
The specific data that are displayed depends on what sensor is connected.
12.12 Simulation
You can simulate values at inputs and outputs for testing purposes:
Current values at current outputs
Measured values at inputs
Relay contact opening or closing
Only current values are simulated. Via the simulation function, it is not possible to
calculate the totalized value for the flow or rainfall.
Before simulation: Enable the inputs and outputs in the Setup menu.
2) Not available for all sensor types.