Diagnostics and troubleshooting Liquiline System CA80SI
128 Endress+Hauser
DIAG/Systemtest
Function Options Info
Temperature Displays the current cuvette heating temperature
Current Displays the current heating current in ampere
Power Displays the heating power in % (current
controller actuating value)
State
Selection
Heating
Heating up
Off
Displays the cuvette heating status
Off for 10 minutes Cuvette heating is switched off for 10 minutes.
Cuvette heating is switched on by default. It can
be switched off temporarily for testing.
On Cuvette heating is switched on.
Magnetic stirrer
Start for 1 minute The magnetic stirrer is switched on for 1 minute.
Stop
Sample pre-heating
Temperature Displays the current sample pre-heating
temperature
Power Displays the heating power in % (current
controller actuating value)
State Displays the current sample pre-heating status
Off for 10 minutes Sample pre-heating is switched off for 10
minutes.
On Sample pre-heating is switched on.
Valves
Valve selection
Selection
None
Reagent RB
Reagent RK
Reagent RN
Open Opens the selected valve.
Close Closes the selected valve.
Pump
Start for 1 minute The pump is switched on for 1 minute.
Stop
Sample channel
Current channel Displays the active measuring channel
Flow rate Displays the current flow of the active measuring
channel
Test channel Select the measuring channel to be tested
After making the selection, press Confirm
Confirm
Dispensers