Liquiline System CA80SI Technical data
Endress+Hauser 151
Connecting optional
modules
With extension modules you can purchase additional functions for your device.
NOTICE
Unacceptable hardware combinations (due to conflicts in power supply)
Incorrect measurements or total failure of the measuring point as a result of heat build-up
or overloading
‣
Find out if the planned upgrade of your device results in a permitted hardware
combination (configurator on www.endress.com/CA80SI).
‣
A maximum of four sensor inputs is permitted.
‣
A maximum of four relay outputs is permitted.
‣
A maximum of eight current inputs and six current outputs is permitted.
‣
A maximum of two "DIO" modules is permitted.
‣
Please contact your Endress+Hauser sales center should you have any questions.
The number of modules is limited by the number of existing cable entries.
Cable entries and possible cable diameters→ 150