Liquiline System CA80SI Installation
Endress+Hauser 23
1-/2-channel device: Installation area for pressure relief valve and filter
A < 1 m
(3.28 !)
A < 1 m
(3.28 !)
L < 3 m
(9.84 !)
L < 0,5 m
(1.64 !)
A0036573
 22 Permitted installation area, engineering unit m (ft)
1
3
4
5
2
A0036671
 23 Installing angle brackets for pressure relief valve and filter
1 Pressure relief valve
2 Angle bracket for pressure relief valve
3 Hose piece (polyurethane hose, length should be < 0.5 m (1.64 ft))
4 Filter
5 Angle bracket for filter
2
1
A0036743
1 Correct flow direction of pressure relief valve (indicated by triangle on pressure relief valve)
2 Correct flow direction of filter (indicated by arrow on filter)
1. Cut hose piece (polyurethane hose) to the required length (< 0.5 m (1.64 ft)).
2. Mount pressure relief valve in angle bracket: Unscrew coupling nut, guide pressure
relief valve through round opening, screw coupling nut back on.
3. Attach the hose piece to the push-in connector of the pressure relief valve.