Commissioning Liquiline System CA80SI
52 Endress+Hauser
Visual inspection of the liquid-bearing lines
Have the bottles with reagents, and standard been inserted and connected?
Is the magnetic stir bar lying flat in the measuring chamber?
10.5.1 Starting the measurement
1. Select start condition Immediate under Menu/Setup/Analyzer/Measurement/
Start condition/Immediate. The analyzer starts immediately with the measurement
cycle, once the system changes to automatic mode.
2. If necessary, adjust the measuring interval under Menu/Setup/Analyzer/
Measurement/Measuring interval.
3. If necessary, adjust the calibration interval under Menu/Setup/Analyzer/
Calibration/Calibration interval.
4. If necessary, adjust the sequence of sample channels under Menu/Setup/Analyzer/
Measurement/Measuring interval/Sequence of measurements.
5. Start automatic mode: Press the MODE soft key and select Start automatic mode .
The display shows Current mode- Automatic.
Re-attach the cover in front of the cuvette assembly.