Commissioning Liquiline System CA80SI
52 Endress+Hauser
Visual inspection of the liquid-bearing lines
‣
Have the bottles with reagents, and standard been inserted and connected?
‣
Is the magnetic stir bar lying flat in the measuring chamber?
10.5.1 Starting the measurement
1. Select start condition Immediate under Menu/Setup/Analyzer/Measurement/
Start condition/Immediate. The analyzer starts immediately with the measurement
cycle, once the system changes to automatic mode.
2. If necessary, adjust the measuring interval under Menu/Setup/Analyzer/
Measurement/Measuring interval.
3. If necessary, adjust the calibration interval under Menu/Setup/Analyzer/
Calibration/Calibration interval.
4. If necessary, adjust the sequence of sample channels under Menu/Setup/Analyzer/
Measurement/Measuring interval/Sequence of measurements.
5. Start automatic mode: Press the MODE soft key and select Start automatic mode .
The display shows Current mode- Automatic.
‣
Re-attach the cover in front of the cuvette assembly.