Severity of the
event
Default diagnostic category Source of the
event
Bit Events within the group
Configuration 21 S442: Frequency output
S443: Pulse output
Process 20 S801: Energy too low
S825: Operating temperature
S921: Change of reference
S942: In safety distance
1)
S943: In blocking distance
S944: Level range
S968: Level limited
1) This event can be removed from the group in order to define its behavior individually; see section
"Configurable area".
Severity of the
event
Default diagnostic category Source of the
event
Bit Events within the group
Lowest severity Maintenance required (M) Sensor 19 not used in Micropilot
Elektronics 18 M272: Main electronics failure
M311: Electronics failure
Configuration 17 M438: Data set
Process 16 M950: Advanced diagnostic
occurred