Commissioning Proline Promass E 300 PROFINET
114 Endress+Hauser
Parameter Prerequisite Description Selection / User
interface / User
entry
Factory setting
Switch-off value In the Relay output function
parameter, the Limit option is
selected.
Enter measured value for the
switch-off point.
Signed floating-point
number
Country-specific:
• 0 kg/h
• 0 lb/min
Switch-off delay In the Relay output function
parameter, the Limit option is
selected.
Define delay for the switch-off
of status output.
0.0 to 100.0 s –
Switch-on value The Limit option is selected in
the Relay output function
parameter.
Enter measured value for the
switch-on point.
Signed floating-point
number
Country-specific:
• 0 kg/h
• 0 lb/min
Switch-on delay In the Relay output function
parameter, the Limit option is
selected.
Define delay for the switch-on
of status output.
0.0 to 100.0 s –
Failure mode – Define output behavior in
alarm condition.
• Actual status
• Open
• Closed
–
* Visibility depends on order options or device settings
10.5.11 Configuring the local display
The Display wizard guides you systematically through all the parameters that can
configured for configuring the local display.
Navigation
"Setup" menu → Display
‣
Display
Format display
→ 115
Value 1 display
→ 116
0% bargraph value 1
→ 117
100% bargraph value 1
→ 117
Value 2 display
→ 117
Value 3 display
→ 117
0% bargraph value 3
→ 117
100% bargraph value 3
→ 117
Value 4 display
→ 117