Commissioning Proline Promass E 300 PROFINET
122 Endress+Hauser
Parameter overview with brief description
Parameter Prerequisite Description Selection / User
interface / User
entry
Factory setting
Corrected volume flow calculation Select reference density for
calculating the corrected
volume flow.
Fixed reference
density
Calculated
reference density
External reference
density
Current input 1
*
Current input 2
*
External reference density In the Corrected volume flow
calculation parameter, the
External reference density
option is selected.
Shows external reference
density.
Floating point
number with sign
Fixed reference density The Fixed reference density
option is selected in the
Corrected volume flow
calculation parameter
parameter.
Enter fixed value for reference
density.
Positive floating-
point number
Reference temperature The Calculated reference
density option is selected in
the Corrected volume flow
calculation parameter
parameter.
Enter reference temperature
for calculating the reference
density.
–273.15 to 99 999 °C Country-specific:
+20 °C
+68 °F
Linear expansion coefficient The Calculated reference
density option is selected in
the Corrected volume flow
calculation parameter
parameter.
Enter linear, medium-specific
expansion coefficient for
calculating the reference
density.
Signed floating-point
number
Square expansion coefficient The Calculated reference
density option is selected in
the Corrected volume flow
calculation parameter
parameter.
For media with a non-linear
expansion pattern: enter the
quadratic, medium-specific
expansion coefficient for
calculating the reference
density.
Signed floating-point
number
* Visibility depends on order options or device settings
10.6.3 Carrying out a sensor adjustment
The Sensor adjustment submenu contains parameters that pertain to the functionality of
the sensor.
Navigation
"Setup" menu → Advanced setup → Sensor adjustment
Sensor adjustment
Installation direction
→  123
Zero point adjustment
→  123