Commissioning Proline Promass E 300 PROFINET
134 Endress+Hauser
Using the parameter to reset the access code
Navigation
"Setup" menu → Advanced setup → Administration → Reset access code
‣
Reset access code
Operating time
→ 134
Reset access code
→ 134
Parameter overview with brief description
Parameter Description User interface / User entry
Operating time Indicates how long the device has been in operation. Days (d), hours (h), minutes (m) and seconds
(s)
Reset access code Reset access code to factory settings.
For a reset code, contact your Endress+Hauser service
organization.
The reset code can only be entered via:
• Web browser
• DeviceCare, FieldCare (via service interface CDI-RJ45)
• Fieldbus
Character string comprising numbers, letters
and special characters
Using the parameter to reset the device
Navigation
"Setup" menu → Advanced setup → Administration
Parameter overview with brief description
Parameter Description Selection
Device reset Reset the device configuration - either entirely or in part - to a
defined state.
• Cancel
• To delivery settings
• Restart device
• Restore S-DAT backup
*
* Visibility depends on order options or device settings
10.7 Simulation
The Simulation submenu enables you to simulate, without a real flow situation, various
process variables in the process and the device alarm mode and to verify downstream
signal chains (switching valves or closed-control loops).
Navigation
"Diagnostics" menu → Simulation
‣
Simulation
Assign simulation process variable
→ 136