Diagnostics and troubleshooting Proline Promass E 300 PROFINET
198 Endress+Hauser
Diagnostic information Remedy instructions
No. Short text
491 Current output 1 to n simulation Deactivate simulation
Measured variable status
Quality Good
Quality substatus Ok
Coding (hex) 0x80 to 0x83
Status signal C
Diagnostic behavior Warning
Influenced measured variables
Diagnostic information Remedy instructions
No. Short text
492 Simulation frequency output 1 to n Deactivate simulation frequency output
Measured variable status
Quality Good
Quality substatus Ok
Coding (hex) 0x80 to 0x83
Status signal C
Diagnostic behavior Warning
Influenced measured variables
Diagnostic information Remedy instructions
No. Short text
493 Simulation pulse output 1 to n Deactivate simulation pulse output
Measured variable status
Quality Good
Quality substatus Ok
Coding (hex) 0x80 to 0x83
Status signal C
Diagnostic behavior Warning
Influenced measured variables