Proline Promass E 300 PROFINET Technical data
Endress+Hauser 231
Status input
Maximum input values DC –3 to 30 V
If status input is active (ON): R
i
>3 kΩ
Response time Configurable: 5 to 200 ms
Input signal level Low signal: DC –3 to +5 V
High signal: DC 12 to 30 V
Assignable functions Off
Reset the individual totalizers separately
Reset all totalizers
Flow override