EasyManua.ls Logo

Endress+Hauser Proline Promass E 300 - Page 235

Endress+Hauser Proline Promass E 300
264 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Proline Promass E 300 PROFINET Technical data
Endress+Hauser 235
0 to 20 mA
Failure mode Choose from:
• Maximum alarm: 22 mA
• Freely definable value between: 0 to 20.5 mA
Pulse/frequency/switch output
Pulse output
Failure mode Choose from:
• Actual value
• No pulses
Frequency output
Failure mode Choose from:
• Actual value
• 0 Hz
• Defined value (f
max
2 to 12 500 Hz)
Switch output
Failure mode Choose from:
• Current status
• Open
• Closed
Relay output
Failure mode Choose from:
• Current status
• Open
• Closed
Local display
Plain text display With information on cause and remedial measures
Backlight Red backlighting indicates a device error.
Status signal as per NAMUR recommendation NE 107
Interface/protocol
• Via digital communication:
PROFINET
• Via service interface
• CDI-RJ45 service interface
• WLAN interface
Plain text display With information on cause and remedial measures
Web browser
Plain text display With information on cause and remedial measures