Endress+Hauser Proline Promass E TB2 HART - 18.6 Performance characteristics

Endress+Hauser Proline Promass E TB2 HART
116 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Technical data Proline Promass E TB2 HART
100 Endress+Hauser
18.6 Performance characteristics
Reference operating
conditions
Error limits following ISO/DIS 11631
Water with +20 to +30 °C (+68 to +86 °F) at 2 to 4 bar (29 to 58 psi)
Specifications as per calibration protocol ±5 °C (±9 °F) and ±2 bar (±29 psi)
Accuracy based on accredited calibration rigs that are traced to ISO 17025
To obtain measured errors, use the Applicator sizing tool(® ä 94)
Maximum measured error
In addition to the values indicated, the measured error at the current output is typically ±4 mA.
o.r. = of reading; 1 g/cm
3
= 1 kg/l; T = fluid temperature
Mass flow and volume flow (liquids)
±0.25 % o.r.
Mass flow (gases)
±0.75 % o.r.
Design fundamentals (® ä 102)
Density (liquids)
±0.0005 g/cm
3
(under reference conditions)
±0.0005 g/cm
3
(after field density calibration under process conditions)
±0.02 g/cm
3
(over the entire measuring range of the sensor)
Temperature
±0.5 °C ± 0.005 × T °C (±0.9 °F ± 0.005 × (T – 32) °F)
Zero point stability
DN
Zero point stability
[mm] [in] [kg/h] or [l/h] [lb/min]
8 3/8 0.24 0.0088
15 ½ 0.78 0.0287
25 1 2.16 0.0794
40 5.40 0.1985
50 2 8.40 0.3087