Proline Promass E TB2 HART Diagnostics and troubleshooting
Endress+Hauser 85
Diagnostics
/../Diagnoselist
Diagnostics3
Diagnostics2
F273Mainelectronic
A0014006-EN
Calling up and closing the remedial measures
1. Press F.
Ã
The message for the remedial measures for the selected diagnostics event opens.
2. Press S + O simultaneously.
Ã
The message about the remedial measures closes.
About the structure of the remedial measure message (® ä 83)
12.5 Overview of diagnostics events
Each diagnostics event is assigned a certain event level at the factory. The user can change this
assignment for certain diagnostics events.
12.5.1 Adapting event level
You can change the event level of a few diagnostics events using the Event no. xxx parameter.
Navigation path
"Expert" menu ® System ® Diagn. handling ® Event level ® Event no. xxx
0723-1
Eventno.044
/../Eventlevel
Eventno.801
Eventno.274
Warning
A0014048-EN
You can assign the following options to the diagnostics event as the event level:
Event level
Description
Alarm The measurement is interrupted. The signal outputs take on the defined alarm condition. A
diagnostic message is generated.
Warning The device continues to measure. A diagnostic message is generated.
Logbook entry only The device continues to measure. The diagnostic message is entered in the Event logbook (events
list) submenu only and not displayed in alternation with the measured value display.
Ignore The diagnostics event is ignored, and no diagnostic message is generated or entered.
12.5.2 Diagnostics for the sensor
Diagnostics event
Event text Remedial measures
Event level from the
factory
F022 Sensor temperature Temperature circuit failure!
1. Change main electronic module.
2. Change sensor.