10.4.6 Configuring the partial filled pipe detection
The Partial filled pipe detection wizard guides you systematically through all parameters that
have to be set for configuring the monitoring of the pipe filling.
Navigation path
"Setup" menu ® Partial filled pipe detection
Sequence of the wizard
Assignprocess
variable
Off Massflow Volumeflow
Corrected
volumeflow
Highvalue
partialfilled
pipedetection
Lowvalue
partialfilled
pipedetection
Responsetime
part.filled
pipedetect.
Endofwizard
A0013801-EN
å 17 "Partial filled pipe detection" wizard in the "Setup" menu
Parameter overview with brief description
Parameters
Description Selection/
User entry
Factory
setting
Assign process variable Select a process variable to detect
empty or partially filled pipes.
• Off
• Density
• Reference density
Density
Low value partial filled pipe
detection
Enter a lower limit value to
activate detection of an empty or
partially filled pipe.
Max. 15-digit, positive floating-
point number
Country-
dependent:
• 0.2 kg/l
• 12.5 lb/cf
High value partial filled pipe
detection
Enter an upper limit value to
activate detection of an empty or
partially filled pipe.
Max. 15-digit, positive floating-
point number
Country-
dependent:
• 6 kg/l
• 374.6 lb/cf
Response time part. filled pipe
detect.
Enter the time interval until the
diagnostics message S862
Partly filled pipe detection is
displayed for an empty or partially
filled pipe.
0 to 100 s 1 s