Proline Promass E TB2 HART Commissioning
Endress+Hauser 65
10.4.6 Configuring the partial filled pipe detection
The Partial filled pipe detection wizard guides you systematically through all parameters that
have to be set for configuring the monitoring of the pipe filling.
Navigation path
"Setup" menu ® Partial filled pipe detection
Sequence of the wizard
Assignprocess
variable
Off Massflow Volumeflow
Corrected
volumeflow
Highvalue
partialfilled
pipedetection
Lowvalue
partialfilled
pipedetection
Responsetime
part.filled
pipedetect.
Endofwizard
A0013801-EN
å 17 "Partial filled pipe detection" wizard in the "Setup" menu
Parameter overview with brief description
Parameters
Description Selection/
User entry
Factory
setting
Assign process variable Select a process variable to detect
empty or partially filled pipes.
Off
Density
Reference density
Density
Low value partial filled pipe
detection
Enter a lower limit value to
activate detection of an empty or
partially filled pipe.
Max. 15-digit, positive floating-
point number
Country-
dependent:
0.2 kg/l
12.5 lb/cf
High value partial filled pipe
detection
Enter an upper limit value to
activate detection of an empty or
partially filled pipe.
Max. 15-digit, positive floating-
point number
Country-
dependent:
6 kg/l
374.6 lb/cf
Response time part. filled pipe
detect.
Enter the time interval until the
diagnostics message S862
Partly filled pipe detection is
displayed for an empty or partially
filled pipe.
0 to 100 s 1 s