EasyManua.ls Logo

Endress+Hauser Proline Promass E TB2 HART - Configuring the Totalizer

Endress+Hauser Proline Promass E TB2 HART
116 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Commissioning Proline Promass E TB2 HART
70 Endress+Hauser
10.5.4 Configuring the totalizer
You can configure each totalizer in the three submenus Totalizer 1-3.
Navigation path
"Setup" menu ® Advanced setup ® Totalizer 1
"Setup" menu ® Advanced setup ® Totalizer 2
"Setup" menu ® Advanced setup ® Totalizer 3
Structure of the submenu
Totalizer 1-3
®
Assign process variable
Unit
Operating mode totalizer 1-3
Failure mode
Parameter overview with brief description
Parameters
Prerequisite Description Selection/
User entry
Factory setting
Assign process
variable
- Select the process
variable for totalizer.
Result
The selection
determines the
picklist of the Unit
parameter.
Off
Mass flow
Volume flow
Corrected
volume flow
Mass flow
Unit One of the following options is
selected in the Assign process
variable parameter:
Mass flow
Volume flow
Corrected volume flow
Select the process
variable of the
totalizer.
Unit picklist Country-dependent:
kg
lb
Operation mode
totalizer 1-3
One of the following options is
selected in the Assign process
variable parameter:
Mass flow
Volume flow
Corrected volume flow
Define how the
totalizer totalizes
the flow.
Net flow total
Forward flow
total
Reverse flow total
Net flow total
Failure mode One of the following options is
selected in the Assign process
variable parameter:
Mass flow
Volume flow
Corrected volume flow
Define how the
totalizer behaves in
an alarm condition.
Stop
Actual value
Last valid value
Stop

Table of Contents