Proline Promass E TB2 HART Commissioning
Endress+Hauser 69
Parameters Description Selection/
User entry
Factory setting
Reference density unit Select the unit for reference
density.
Result
The selected unit applies for:
Current outputs
Low value partial filled
pipe detection
High value partial filled
pipe detection
Simulation process
variable
Fixed reference density
Density adjustment (in the
Expert menu)
Unit picklist
Country-dependent:
kg/Nl
lb/scf
Temperature unit Select the unit for
temperature.
Result
The selected unit applies for:
Current outputs
Reference temperature
Simulation process
variable
Unit picklist Country-dependent:
°C (Celsius)
°F (Fahrenheit)
Length unit Select the unit of length for
the nominal diameter.
Unit picklist Country-dependent:
mm
in
Pressure unit Select the unit for pipe
pressure.
Unit picklist Country-dependent:
bar a
psi a
10.5.3 Carrying out a sensor adjustment
The Sensor adjustment submenu contains parameters that pertain to the functionality of the
sensor.
Navigation path
"Setup" menu ® Advanced setup ® Sensor adjustment
Structure of the submenu
Sensor adjustment
®
Installation direction
Zero point adjustment control
Parameter overview with brief description
Parameters
Description Selection/
User entry
Factory setting
Installation direction Change the sign of the
direction of flow of the fluid.
Flow in arrow direction
Flow against arrow
direction
Flow in arrow direction
Zero point adjustment
control
Start the zero point
adjustment.
Cancel
Start
Cancel

Table of Contents