Commissioning Proline Promass E TB2 HART
68 Endress+Hauser
Pressure unit
Parameter overview with brief description
Parameters Description Selection/
User entry
Factory setting
Mass flow unit Select the unit for mass flow.
Result
The selected unit applies for:
Current outputs
Low flow cut off
Simulation process
variable
Unit picklist Country-dependent:
kg/h
lb/min
Mass unit Select the unit for mass.
Result
The selected unit is taken
from:
Mass flow unit
Unit picklist Country-dependent:
kg
lb
Volume flow unit Select the unit for volume
flow.
Result
The selected unit applies for:
Current outputs
Low flow cut off
Simulation process
variable
Unit picklist Country-dependent:
l/h
gal/min (us)
Volume Select the unit for volume.
Result
The selected unit is taken
from:
Volume flow unit
Country-dependent
l
gal (us)
Corrected volume flow unit Select the unit for corrected
volume flow.
Result
The selected unit applies for:
Current outputs
Low flow cut off
Simulation process
variable
Unit picklist Country-dependent:
Nl/h
Scf/min
Corrected volume unit Select the unit for standard
volume.
Result
The selected unit is taken
from:
Corrected volume flow unit
Unit picklist Country-dependent:
Nl
Scf
Density unit Select the unit for density.
Result
The selected unit applies for:
Current outputs
Low value partial filled
pipe detection
High value partial filled
pipe detection
Simulation process
variable
Density adjustment (in the
Expert menu)
Unit picklist Country-dependent
kg/l
lb/cf