Commissioning Proline Promass E TB2 HART
64 Endress+Hauser
10.4.5 Configuring the low flow cut off
The Low flow cut off wizard guides you systematically through all parameters that have to be set
for configuring the low flow cut off.
Navigation path
"Setup" menu ® Low flow cut off
Sequence of the wizard
Assignprocess
variable
Off Massflow Volumeflow
Correctedvolume
flow
Offvalue
lowflow
cutoff
Onvalue
lowflow
cutoff
Pressureshock
suppression
Endofwizard
A0013799-EN
å 16 "Low flow cut off" wizard in the "Setup" menu
Parameter overview with brief description
Parameters
Description Selection/
User entry
Factory setting
Assign process variable Select the process variable for
low flow cut off.
• Off
• Mass flow
• Volume flow
• Corrected volume flow
Mass flow
On value low flow cut off Enter the on-value for the
low flow cut off.
Max. 15-digit, positive
floating-point number
For liquids: depends on
country and nominal
diameter
Off value low flow cut off Enter the off-value for the
low flow cut off.
0 to 100 % 50 %
Pressure shock suppression Enter the time interval for
signal suppression (= active
pressure shock suppression).
0 to 100 s 0 s