Proline Prowirl 73 Description of device functions
Endress+Hauser 123
END VALUE
FREQUENCY
Prerequisite
The function is only available if FREQUENCY was selected in the OPERATION MODE
function (ä 122).
Description
Use this function to specify an end value frequency for the frequency output. The
associated measured value of the measuring range is specified in the VALUE f HIGH
function (ä 123).
The output signal is symmetrical in the FREQUENCY operating mode
(on/off ratio = 1:1).
Example:
End frequency = 1000 Hz, VALUE f HIGH. = 1000 kg/h: i.e. a frequency of 1000 Hz is
output at a flow of 1000 kg/h.
End frequency = 1000 Hz, VALUE f HIGH. = 3600 kg/h: i.e. a frequency of 1000 Hz is
output at a flow of 3600 kg/h.
User input
5-digit fixed point number: 2 to 1000 Hz
Factory setting
1000 Hz
VALUE f LOW Prerequisite
The function is only available if FREQUENCY was selected in the OPERATION MODE
function (ä 122).
Description
Use this function to assign a value to the start frequency.
The value entered here must be smaller than the value assigned to the VALUE f HIGH
(ä 123). A negative value is only permitted if TEMPERATURE is selected in the
ASSIGN FREQUENCY function (ä 122). The desired span is defined by specifying
the VALUE f LOW and VALUE f HIGH. The appropriate unit is taken from the SYSTEM
UNITS (ä 102) or MEASURING VALUES (VELOCITY) group.
User input
5-digit floating-point number
Factory setting
Depends on the option selected in the ASSIGN FREQUENCY function.
0 [UNIT VOLUME FLOW]
–0 °C (converted to the UNIT TEMPERATURE)
0 [UNIT MASS FLOW]]
0 [UNIT CORRECTED VOLUME FLOW]]
0 [UNIT HEAT FLOW]
–0 [VELOCITY]
0 [UNIT PRESSURE]
VALUE f HIGH Prerequisite
The function is only available if FREQUENCY was selected in the OPERATION MODE
function (ä 122).
Description
Use this function to assign a value to the end frequency.
The value entered here must be larger than the value assigned to the VALUE f LOW
(ä 123). A negative value is only permitted if TEMPERATURE is selected in the
ASSIGN FREQUENCY function (ä 122). The desired span is defined by specifying
the VALUE f LOW and VALUE f HIGH. The appropriate unit is taken from the SYSTEM
UNITS (ä 102) or MEASURING VALUES (VELOCITY) group.
User input
5-digit floating-point number
Factory setting
Depends on the option selected in the ASSIGN FREQUENCY function.
10 l/s (converted to the UNIT VOLUME FLOW)
200 °C (converted to the UNIT TEMPERATURE)
10 kg/h (converted to the UNIT MASS FLOW)
10 Nm³/h (converted to the UNIT CORRECTED VOLUME FLOW)
10 kW (converted to the UNIT HEAT FLOW)
10 m/s (converted to the unit of VELOCITY)
10 bara (converted to the UNIT PRESSURE)