Description of device functions Proline Prowirl 73
142 Endress+Hauser
OFF-VALUE LOW
FLOW CUT OFF
Description
Use this function to enter the off-value for low flow cut off. Enter the off-value as a
positive hysteresis from the on-value.
A0001245
Fig. 34: Example for the behavior of low flow cutoff
Q Flow rate [volume/time]
tTime
aON-VALUE LOW FLOW CUTOFF = 20 m³/h
b OFF-VALUE LOW FLOW CUTOFF = 10%
c Low flow cutoff active
1 Low flow cutoff is switched on at 20 m³/h
2 Low flow cutoff is switched off at 22 m³/h
HHysteresis
User input
Integer 0 to 100%
Factory setting
50%
VELOCITY WARNING Description
Use this function to activate flow velocity supervision (ON).
If the fluid velocity exceeds the value entered in the LIMIT VELOCITY function
(ä 142) the device issues the notice message "#421 FLOW RANGE" (ä 68).
Options
OFF (function switched off)
ON
Factory setting
OFF
LIMIT VELOCITY Description
Enter the maximum allowable flow velocity (limit velocity).
By switching on the VELOCITY WARNING function (ä 142), a warning message is
output once the limit velocity is overshot.
Unit on the display depends on UNIT LENGTH (ä 106).
User input
5-digit floating-point number
Factory setting
75 m/s
Function description, PROCESS PARAMETER group