Description of device functions Proline Prowirl 73
164 Endress+Hauser
MOLE % C2H6 Description
Use this function to enter the ethane mole-% in the expected natural gas mixture
(max. 10%; "extended range" of the application: max. 20 %).
User input
5-digit floating-point number
Factory setting
0 %
MOLE % C3H8 Description
Use this function to enter the propane mole-% in the expected natural gas mixture
(max. 3.5%; "extended range" of the application: max. 5 %).
User input
5-digit floating-point number
Factory setting
0 %
MOLE % H2O Description
Use this function to enter the water vapor mole-% in the expected natural gas mixture
(max. 0.015 %).
User input
5-digit floating-point number
Factory setting
0 %
MOLE % H2S Description
Use this function to enter the water vapor mole-% in the expected natural gas mixture
(max. 0.02 %).
User input
5-digit floating-point number
Factory setting
0 %
MOLE % H2 Description
Use this function to enter the hydrogen mole-% in the expected natural gas mixture
(max. 10 %).
User input
5-digit floating-point number
Factory setting
0 %
MOLE % CO Description
Use this function to enter the carbon monoxide mole-% in the expected natural gas
mixture (max. 3 %).
User input
5-digit floating-point number
Factory setting
0 %
MOLE % O2 Description
Use this function to enter the oxygen mole-% in the expected natural gas mixture
(max. 0.02 %).
User input
5-digit floating-point number
Factory setting
0 %
Function description, NG AGA8-DC92/ISO 12213-2 group