Waterpilot FMX21 Description of device parameters
Endress+Hauser 91
Prerequisite • "Level selection" = In pressure
• "Calibration mode" = Dry -> entry
• "Calibration mode" = Wet -> display
Factory setting URL of the sensor module
Full height
Write permission Operators/Service engineers/Expert
Description Enter the height value for the upper calibration point (vessel full). The unit is selected via
the "Height unit" parameter.
Prerequisite • "Level selection" = "In height"
• "Calibration mode" = Dry -> entry
• "Calibration mode" = Wet -> display
Factory setting URL is converted to a level unit
Density unit
Write permission Service engineers/Expert
Description Displays the density unit. The measured pressure is converted to a height using the "Height
unit", "Adjust density" and "Process density" parameters.
Options • g/cm
3
• kg/m
3
• kg/dm
3
• lb/in
3
• lb/ft
3
Factory setting g/cm
3
Adjust density
Write permission Operators/Service engineers/Expert
Description Enter the density of the medium used to perform the calibration. The measured pressure is
converted to a height using the "Height unit" and "Adjust density" parameters.
Input: Auto dens. corr. = Off
Display: Auto dens. corr. ≠ Off
Factory setting 1.0