A.2 Technical Specifications - Probes
Process gas temperature:
up to 500 °C ( 932 °F)
up to 1600 °C (2912 °F) with cooling tube
Insertion depth:
KES-6001: 470 mm
KES-6002: 930 mm
KES-6003: 1850 mm
Insertion depth with cooling tube:
500/1000 mm
others on request
Measuring principle:
Zirconium oxide for O
2
MXP for CO
e
Flue gas pressure:
±50 mbar (±0.725 PSIG) to atmospheric pressure
Flue gas flow velocity:
0 to 10 m/s, others on request
Ambient temperature:
-40 °C to +80 °C (-40 °F to +176 °F)
Reaction time (O
2
):
< 1 s (with test gas)
T90 (O
2
):
< 5 s (with test gas)
Probe material:
stainless steel (SS316)
Protection class:
IP65
Detection limit:
< 10 ppm O
2
Power supply:
Through electronic unit
Probe dimensions:
on request