PLATINOUS SERIES CONTROLLER MANUAL
11
2.5. Detail Monitor: Program Monitor
This screen can only be displayed during programmed operations. It contains information on the
program being run.
1. PGM Name
Displays the No. and name of the program being run.
2. Current Step
Displays the No. of the step currently being executed.
3. End Step
Displays the No. of the last step in the program.
4. Temp. SP
Indicated the current target temperature of the step being run. Displays “AIR” below “Temp.” if
air control is active, shows “PRODUCT” below “Temp.” if product control is active. (For details
on air/product temp control, see section 5.11.11)
5. PV
Displays the monitored temperature.
6. Hum. SP
Indicates the current target humidity of the step being run.
(For temperature & humidity chambers only)
7. PV
Displays the monitored humidity.
(For temperature & humidity chambers only)
8. Time to Step End
Displays the time left until the step ends. Time is counted down.
9. Scheduled Test End
Displays the time scheduled for the test to end. The following are not included in the schedule.
Temperature stabilization time when soak control is ON
Time gained when jumping to the next step in programmed operations
10. Total Run Time
1
3
8
9
10
11
12
2
4
5
6
7
12

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ESPEC SCP-220 and is the answer not in the manual?

ESPEC SCP-220 Specifications

General IconGeneral
BrandESPEC
ModelSCP-220
CategoryController
LanguageEnglish

Related product manuals