MT100 Multipoint Flow Meter INSTALLATION
4-20 mA Input
The MT100 is provided with a 4-20 mA input on J8. The connector accepts 28-16 AWG (0.14 mm
2
-1.5 mm
2
) wire (refer to Table 2, page
23 for wire size vs. length info). Connect the external current loop input to J8-5 (IN) and J8-6 (RTN). The 4-20 mA input is used for the
following functions:
● External Input Flow Adjust
● External MT100 Flow Input
● External Control Group Switching
Refer to Extended Operation Modes, page 72 (OPERATION section) for extended operation mode details.
Service Port Connection, USB & Ethernet
Listed below are the MT100 service ports, which are used to configure/monitor the instrument via a PC. See also Configuring the MT100, page 34.
● USB 2.0 – USB Type B connector J22: Use the USB port for local host PC connection to the instrument.
● Ethernet (100Base-T) – modular RJ-45 jack J23: Use the Ethernet port for remote applications in which the host PC communicates with
the instrument over an Ethernet (100Base-T) network
Fluid Components International LLC 31