EasyManua.ls Logo

FCI MT100 - Table 9 - MT100 HART Device Specific Command Groupings; Table 10 - HART Device Specific Commands

FCI MT100
160 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
OPERATION MT100 Multipoint Flow Meter
MT100 HART Device Specific Commands
The MT100 Manufacturer Specific or Device Specific commands start at command 137. Use the device specific commands to setup and
configure the MT100 Series instrument via HART. The MT100 device specific commands are grouped in functional categories as
summarized in Table 9 below.
Table 9 – MT100 HART Device Specific Command Groupings
Group No.
Description
Command Numbers
Group 1
Commands to set up and configure the instrument.
137, 138, 139, 140, 145, 146,
148, 149, 150, 159
Group 2
Commands to set up 4–20 mA output channels including the OUTZ, and
OUTF parameters.
160, 161, 163, 164, 166, 167
Group 3
Commands to view individual FE process. The view is a snapshot of
sensor data at the time of the request; i.e., it does not update.
170, 172, 174, 176
Group 4
Commands to display the factory-set calibrated limit of the instrument for
flow, process temperature and pressure variables.
151, 154, 157
Group 5
Other Category – Commands outside the above listed groups.
179, 180, 181, 182, 183, 184,
185,186, 187, 191. 193, 195
Table 10 below summarizes the instrument’s HART Device Specific command set and the data associated with each command.
Table 10 HART Device Specific Commands
Command 137: Read Totalizer And Rollover Values
Byte
Format
Description
Request Data Bytes
None
Response Data Bytes
0–3
Float
Totalizer
4–7
Integer
Rollover
Code
Class
Description
Response Codes
0
Success
No Command-Specific Errors
1–5
Undefined
6
Error
Device-Specific Command Error
7–127
Undefined
Command 138: Read Totalizer State
Byte
Format
Description
Request Data Bytes
None
Response Data Bytes
0
Unsigned-8
Totalizer State: 0 = OFF; 1 = ON
Code
Class
Description
Response Codes
0
Success
No Command-Specific Errors
1–5
Undefined
6
Error
Device-Specific Command Error
7–127
Undefined
Fluid Components International LLC 58

Table of Contents

Related product manuals