MT100 Multipoint Flow Meter OPERATION
MT100 HART Common Practice Commands
The MT100 supports Common Practice commands 35, 40, 42, 44, 45, 46, 50 and 51. Table 8 below summarizes the instrument’s HART
Common Practice command set and the data associated with each command.
Table 8 – HART Common Practice Commands
Command 35: Write Primary Variable (PV) Range Values
PV Upper and Lower Range Values Units Code
PV Upper Range Value (Customer Max. Flow Limit)
PV Lower Range Value (Customer Min. Flow Limit)
PV Upper and Lower Range Values Units Code
No Command-Specific Errors
Lower Range Value Too High
Lower Range Value Too Low
Upper Range Value Too High
Upper Range Value Too Low
Upper and Lower Range Values Out Of Limits
Note: 1. The value returned in the response data bytes reflects the rounded or truncated value actually used by the
device.
Command 40: Enter/Exit Fixed Current Mode
PV Fixed Current Level (mA units); “0” to Exit Fixed Current
No Command-Specific Errors
Passed Parameter Too Large
Passed Parameter Too Small
Loop Current Not Active (Device in Multidrop Mode)
Notes: 1. Specify a value (in mA) to drive Ch. 1 to a particular output value. Specify “0” to exit the fixed current mode.
Command 42: Perform Device Reset (Soft Reset of Flow Meter)
1
Note: 1. Send Command 42 (no data) to reset the instrument. No response is returned due to reboot.
Fluid Components International LLC 55