33
FERITSCOPEĀ® DMP30 Connecting/replacing probe
3. Open the most recently opened batch: .
4. The Reassign probe? menu page appears.
The menu page displays the probe identifiers of the two probes: the old one
previously assigned to the batch, and the connected new probe.
5. Assign the connected probe to the batch: OK.
A message tells you that a calibration still needs to be assigned to the open
batch.
6. What would you like to do?
- Assign a calibration from the connected probe to the open batch: >
select desired calibration > OK.
This completes the process and the Measurement view of the open batch
appears.
- Perform a new calibration for the open batch: > New > follow the cali-
bration routine, for the procedure, see page 63.
This completes the process and the Measurement view of the open batch
appears.
- Do no t assign a calibration to the open batch: OK
This completes the process and the main menu appears.
Please note that a measurement is possible only if a calibration has
been assigned to the batch.