FERITSCOPE® DMP30 Measure
8.4 Before you start
The measuring area and the probe tip on the probe are clean, free of oil, dust,
etc.
The probe is connected to the gauge.
The gauge is acclimatized and the battery are sufficiently charged for the up-
coming measurement.
Refer to chapter "Parameters that influence the Delta-ferrite content measure-
ment", page 44
Note: For measurements in Fe%, the value for ferrite content percentage
(Fe%) is strictly informational and have no reference to traceable standards.
The ferrite percent specifications (Fe%) on the calibration standards were de-
termined with a round robin based on the metallographic counting method and
is no t traceable.
The desired batch is open in the gauge with the correct measurement settings.
In the event of an online data transfer during the measurement, connections
suitable for Tactile Suite are configured.
During the measurement, you can also perform the following actions
in the Measurement view
Delete single/all measured readings or a measured reading lock that has not
been closed: press the key.
View the statistics of the open/last measured reading block: press the BLOCK-
key
Set block closure manually: BLOCK > CLOSE, available only when automatic
block formation is deactivated