FERITSCOPE® DMP30 Settings for measurement
You can choose between automatic (dynamic) and manual (static) acquisition of the
air reference value:
Dynamic (default mode)
Dynamic, automatic acquisition of the air value each time the probe is lifted off the
surface.
Static
Static, no automatic acquisition of the air value each time the probe is lifted off the
surface. The air reference value has to be measured manually at regular intervals.
Useful for measurements in small cavities or pipes where the minimum lift off
distance of the probe cannot be maintained within the cavity/pipe.
Default in Probe
Information about the acquisition method for the air reference value set in the probe
in the factory.
Before you start
The batch for which you want to make the changes is open in the gauge.
To change the acquisition method for the air reference value
1. > Batch Modify >
OK
> More Settings >
OK
> Air Reference >
OK
> set service
code (159) > OK > set desired acquisition method >
OK
Parameter is activated:
2. Exit the menu: , the setting has now been made.
7.9 Assigning a different calibration to the opened batch
Before you start
The batch for which you want to make the change is open in the gauge.
To assign a different calibration to the open batch
> Batch Modify > OK > More Settings > OK > Calibration Assign > OK > select
the desired calibration from the list > OK
A different calibration has now been assigned.
[ ]
[ ]