56
Calibration FERITSCOPE® DMP30
9.4.1 Calibration step Zero – Procedure
Measure on the calibration standard Base.
Required material
Base: calibration standard with very high ferrite content (ca. 100 FN) from the
calibration set.
To perform the calibration step Zero
1. Perform 5 to 10 measurements on the calibration standard Base.
The display screens should be considered only as examples, the gauge default modes are required here
for the measured reading acquisition.
Repeat Steps A to C for the subsequent measurements on the uncoated refer-
ence piece.
2. Finish the calibration step Zero: OK.
3. What would you like to do?
► Finish the calibration: 1 x
The measuring system is now calibrated to the zero point (= Zero), the calibra-
tion is finished, the calibration menu with the calibration list appears in the
display
► Continue the calibration with calibration standard(s): continue with "Calibration
step Std – procedure", 58
A B C
Position the gauge
Measured reading
acquisition
Lift off the probe
≥ 4 x probe
measurement
range
Base
Calibration 003
Zero
x
X
106
105.8
n
3
OK
Calibration 003
Zero
x
X
104
104.8
n
4
OK
Calibration 003
Zero
x
X
104
104.8
n
4
OK