58
Calibration FERITSCOPE® DMP30
9.4.2 Calibration step Stdprocedure
Measure on the calibration standard.
Required material
Calibration standard(s) with the desired ferrite contents from the calibration set
To perform the calibration step Std
1. Perform 5 to 10 measurements on the calibration standard (Std. 1).
2.
The display screens should be considered only as examples, the gauge default modes are required here
for the measured reading acquisition.
Repeat Steps A to C for the subsequent measurements on the uncoated refer-
ence piece.
3. Enter the nominal value of the Std. 1: STD > set nominal value> OK.
4. Finish the calibration step Std 1: OK.
5. What would you like to do?
Finish the calibration: 1 x
The measuring system is now calibrated to a reference value (Std 1), the cali-
bration is finished, the calibration menu with the calibration list appears in the
display
Continue the calibration with the 2nd/3rd calibration standard: OK > repeat
steps 1. to 4. with the second/third calibration standard (Std 2/Std 3)
The measuring system is now calibrated to two/three reference values (Std 1
and Std 2/Std 3), the calibration is finished, the calibration menu with the cali-
bration list appears in the display.
A B C
Position the gauge
Measured reading
acquisition
Lift off the probe
4 x probe
measurement
range
Calibration
standard
Calibration 003
Zero
x
X
0.65
0.642
n
0.64
3
OK
Calibration 003
Zero
x
X
0.64
0.642
n
0.64
4
OK
Calibration 003
Zero
x
X
0.64
0.642
n
0.64
4
OK