FERITSCOPE® DMP30
Calibration
9.14 Deleting a calibration
You can delete individual calibrations from the memory of the probe/DMP-F probe
adapter.
Before you start
The probe is connected to the gauge containing the memory from which you
want to delete a calibration.
In the gauge is the batch open to which a calibration is assigned for the same
measurand (FN or Fe%), as the one you want to delete.
If the calibration you want to delete is locked, you need to know the lock code.
To delete a calibration
1. Select the calibration you want to delete: > Calibrations > OK > select name
of calibration > OK.
2. Prompted to set the lock code?
- Yes: Set the lock code and deactivate the lock function, 70.
Function is deactivated:
- No: Continue with step 3.
3. Select the delete function: Delete > OK.
4. Start the delete process: .
- If the calibration of the opened batch is deleted, you have to confirm the
message: .
5. Exit the menu: , the delete process has finished.
[ ]